Seaward Primetest 250+ PAT Tester

£504.00

Compare


Enabling simple, professional and fast PAT testing, this versatile tester features all of the required tests of the IET 5th Edition Code of Practice along with USB download. It is also fully compatible with the Test n Tag Pro serial printer for quick and easy label printing without any setup required – just plug and print.

Using Seaward’s free datalogger program and pressing a single download key allows test result data to be quickly and easily transferred from the PrimeTest 250+ to PC at the press of a button.

Data is downloaded in a simple CSV spreadsheet format for your PAT records or alternatively it can be imported directly into Seaward’s PATGuard 3 software for more enhanced record keeping, reporting and certification.

Seaward’s unique ‘ZAP’ technology ensures that earth continuity measurements are accurate in situations where other testers struggle to provide reliable measurements, such as measuring through rust, limescale and other surface contamination.

The PrimeTest Series Testers can be used with our free PATMobile App as part of a combined record keeping and reporting solution. The PATMobile App works with the PrimeTest 250+ to enable the input and storage of test records alongside barcode scanning, wireless label printing, photo tagging and mobile reporting.

Built in Britain and used by 1,000’s of organisations every day.

Simple data management

A Seaward datalogger program is supplied as standard and this along with the single key download feature allows measurement data to be quickly and easily transferred from the PrimeTest 250+ to PC at the press of a button; whilst all records are date stamped for complete traceability.

Data is downloaded in CSV format and can be accessed by a wide range of software packages or alternatively imported directly into PATGuard 3 Elite for truly professional reports and certificates.

As with all Seaward battery powered PAT testers, accurate earth continuity tests are a given thanks to the unique patent pending “zap” circuit design.

Plug & Print

The PrimeTest 250+ (403A910) can be used in conjunction with the new Test n Tag Pro serial printer to further speed up the process of PAT testing and labelling. The Test n Tag Pro label rolls provide over 400 high quality labels allowing you to produce professional, low cost inspection and test labels at the press of a button.

PrimeTest 250+ has an internal memory capable of storing upto 999 appliance test records and the single key save allows records to be stored at the press of a button.

Transfer data via a USB cable directly to a PC using the free datalogger software. The data downloads in the industry standard CSV format and can be read directly into spreadsheets or other PC applications.

For those who require professional reports and certificates as well as complete data traceability and integrity, the downloaded data is also fully compatible with Seaward’s PATGuard 3 Elite software; which allows complete test records to be easily managed on a PC – offering total peace of mind and quicker, easier management of scheduling inspections and tests.

Key Features

  • Very easy to use
  • Plug and print – easy label printing with no setup
  • Date stamped results for full traceability
  • Single key save
  • Single key download
  • Download software included
  • Store 999 appliance records
  • Industry standard download format
  • Comprehensive range of electrical tests
  • Test 3 phase appliances

 

Technical Specifications

Earth Continuity
Output Current200mA minimum into 2Ω
Test Voltage (o/c)9VDC nominal
Test Duration (Class I Test)2s nominal
Display Range0.01 – 19.99Ω
Pass Value (Class I Test)0.1Ω/0.2Ω/0.3Ω
Equivalent Leakage
Test Voltage40V Nominal AC 50Hz Open Circuit
Test Duration2s
Display Range0.1 – 19.99mA
Pass Value5mA
Protective Conductor Current
Test VoltageSupply Voltage, max current 13A
Test Duration120s maximum
Display Range0.15 – 9.99mA
Pass Value5mA
3 phase leakage current (Requires 3 phase adaptor accessory)
Test Voltage3 phase supply voltage
Test DurationContinuous measurement
Display Range0.10 – 9.99mA
Pass Value5mA
Power Socket Check
Input Voltage Range225 – 254VAC max current 13A UK
Indicates Configuration of Voltage PotentialLine potential phase to earth. Line potential phase to neutral. Line potential neutral to earth
Insulation Resistance
Test Voltage250/500VDC @1mA nomina
Test Current< 2mA (s/c)
Test Duration (Class I or II)2s
Display Range0.01 – 19.99MΩ
Pass Value Class I1.0MΩ
Pass Value Class II2.0MΩ
IEC Lead Test
Test Voltage5VDC nominal
Test DurationTime for test (2s max)
TestLive / Neutral Checks for o/c, s/c and crossed
Touch Current
Test VoltageSupply voltage, max current 13A
Test Duration120s maximum
Display Range0.10 – 5mA
Pass Value5mA

Down load the PT250+ (5TH ED) DATA SHEET data sheet

Download the SEAWARD 2014 CATALOGUE

Reviews

There are no reviews yet.

Be the first to review “Seaward Primetest 250+ PAT Tester”

Your email address will not be published. Required fields are marked *