Fluke 705 Loop Calibator

£1,035.00

Compare

The Fluke 705 Loop Calibrator is a cost-effective, integrated solution for calibration, repair and maintenance of current loops. It features:

  • mA sourcing, simulation and measurement
  • 0.02% accuracy and 0.001 mA resolution
  • Simultaneous mA and % of span display
  • Selectable step or ramp outputs
  • 24 V Loop supply with mA measure
  • 0-28 VDC measurement to check Loop Voltage
  • Easy battery exchange without opening the case
  • Compact case with separate protective holster
  • Three year warranty
  • A large display and simple, pushbutton interface for ease of use.
  • Simultaneous mA and % readout for quick, easy interpretation of readings.
  • mA accuracy of 0.02%
  • Push button 25% steps for fast, easy linearity checks. Save time over traditional knob type controls.
  • “Span Check” for fast confirmation of zero and span.
  • Selectable slow ramp, fast ramp, and step to provide smooth outputs for valve slewing and loop functional tests.
  • 24 volt internal loop supply, so you can power and read a transmitter at the same time without carrying a DMM.
  • Single 9V battery that is easily changed.
  • 0 – 20 mA or 4 – 20 mA default start up modes.

 

Specifications
Current Measurement
Range: 0 to 24 mA
Resolution: 0.001 mA
Accuracy: 0.02% Rdg + 2 LSD
Current Sourcing
Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.025% Rdg + 2 LSD
Drive Capability: 1000 Ω @ 24 mA
Current Simulation
Range: 0 to 20 mA or 4 to 20 mA
Accuracy: 0.025% Rdg + 2 LSD
Ext. loop voltage requirement: 12-30 V
Loop Power
 While measuring mA, 24 V
Voltage measurement
Range: 0 to 28 V DC
Resolution: 1 mV
Accuracy: 0.025% Rdg + 2 LSD
Environmental Specifications
Operating Temperature
 -10 ºC to 55 ºC
Storage Temperature
 -40 ºC to 60 ºC
Humidity (Without Condensation)
 0 to 90% (0 ºC to 35 ºC); 0 to 70% (35 ºC to 55 ºC)
Temperature Coefficient
 -10 ºC to 18 ºC, 28 ºC to 55 ºC, ±.005% of range per ºC
Safety Specifications
Safety
 CSA C22.2 No. 1010.1:1992
EMC
 EN50082-1:1992 and EN55022:1994 Class B
Maximum voltage
 30 V
Mechanical & General Specifications
Size
 152 x 70 x 41 mm
Weight
 340g
Batteries
 One 9V battery
Warranty
 Three years
Shock & Vibration
 per MIL-T-28800 for a Class 2 Instrument
Display current and % of span
 Yes
Auto step, auto ramp
 Yes
Span Check
 Yes

The Fluke 705 loop calibrator is supplied with the following;

  • AC70A Alligator Clips
  • C10 protective yellow holster with Flex-Stand™
  • Instruction sheet
  • TL75 safety-designed test lead set
  • NIST- traceable calibration report and data
  • Free ACT calibration certificate

Click on the links below for further information;

FLUKE 705 TECH DATA

Select model

30G, 150G, 300G

Reviews

There are no reviews yet.

Be the first to review “Fluke 705 Loop Calibator”

Your email address will not be published. Required fields are marked *

Select at least 2 products
to compare